Light and Sound ยป Practical: Measuring Sound Frequency

What you'll learn this session

Study time: 30 minutes

Edexcel spec: 3.27P

  • How to set up a microphone and an oscilloscope to measure the frequency of a sound
  • How to use the time base setting to find the time period from a trace
  • How to calculate the frequency from the time period
  • How to make the result more accurate and check it

๐Ÿ”’ Unlock Full Course Content

Sign up to access the complete lesson and track your progress!

Unlock This Course

The idea of the practical

This content is examined in Paper 2 only.

In Oscilloscopes, Pitch and Loudness you saw that more waves across the screen means a higher frequency. This practical goes one step further: you use the trace to measure the frequency in hertz.

The plan is simple. Time runs across the oscilloscope screen, so the width of one complete wave tells you the time period of the sound. Once you have the time period, the frequency comes from frequency = 1 / time period.

Key terms:

  • Division: one square of the grid on the oscilloscope screen.
  • Time base: the oscilloscope setting that fixes how much time each division across the screen stands for, for example 2 ms per division (2 ms/div).
  • Gain: the oscilloscope setting that fixes how many volts each division up the screen stands for. It changes the height of the trace.
  • Tuning fork: a metal fork with two prongs that vibrates at one fixed frequency when struck. The frequency is marked on it.
  • Signal generator: a device that produces an electrical signal of a frequency you choose with a dial. Connected to a loudspeaker, it gives a steady note.

Apparatus

You need a microphone connected by a cable to the input of an oscilloscope, and a source of a steady, pure note. There are two common choices.

🎵 Tuning fork

Simple, and its frequency is marked on it, so you can check your answer. But the sound is quiet and fades in a few seconds, so the trace shrinks while you are reading it.

🔊 Signal generator and loudspeaker

The note stays the same for as long as you like, so the trace stays steady. You can also change the frequency with the dial and test several notes.

To draw the apparatus in the exam, draw the source on the left (a loudspeaker joined to a box labelled 'signal generator', or a tuning fork). Draw a microphone a short distance in front of it. Then draw a wire from the microphone to a box with a screen, labelled 'oscilloscope'.

Method

  1. Connect the microphone to the input of the oscilloscope and switch the oscilloscope on. With no sound, check that the trace is a flat line on the zero line across the middle of the screen.
  2. Make the sound. Strike the tuning fork on a cork or rubber block and hold the prongs close to the microphone. Or switch on the signal generator and loudspeaker and place the microphone in front of the loudspeaker.
  3. Adjust the gain until the trace is as tall as possible while still fitting on the screen.
  4. Adjust the time base until the trace is steady and shows a few complete waves across the screen.
  5. If the oscilloscope has a button that freezes the trace, press it. This is very helpful with a tuning fork.
  6. Count the number of divisions across the screen for one complete wave, for example from one crest to the next crest. Better still, count the divisions for several complete waves.
  7. Write down the time base setting.
  8. Calculate the time period, then the frequency.
  9. Repeat and find a mean. With a signal generator, repeat for other frequencies.

Why adjust the settings? If the waves are squashed together or the trace is tiny, you cannot read the divisions accurately. A tall trace with a few wide waves is the easiest to measure.

From the trace to the frequency

There are two steps.

Step 1: time period = number of divisions for one complete wave × time base setting

Step 2: frequency = 1 / time period, f = 1 / T

Remember, f = 1 / T is given on the formulae page, and it only gives hertz if the time period is in seconds. Time base settings are usually in milliseconds, so divide by 1000 before step 2.

Worked example 1: one wave

A signal generator and loudspeaker produce a note. On the oscilloscope, one complete wave is 8 divisions wide. The time base is set to 0.2 ms/div. Calculate the frequency.

time period = number of divisions × time base = 8 × 0.2 = 1.6 ms

1.6 ms = 1.6 ÷ 1000 = 0.0016 s

f = 1 / T = 1 ÷ 0.0016 = 625 Hz

Measuring across several waves

You can only read the screen to about the nearest 0.1 of a division. If one wave is only 2 divisions wide, an error of 0.1 of a division is a large fraction of your reading. If you measure across several waves instead, the same small reading error is shared out between all of them, so the error in the time for one wave is much smaller.

Count carefully. From the first crest to the sixth crest is 5 complete waves, not 6. Call the first crest zero.

Worked example 2: several waves and a tuning fork

A tuning fork is marked 256 Hz. On the frozen trace, 5 complete waves take up 9.8 divisions. The time base is set to 2 ms/div. Calculate the frequency.

time for 5 waves = 9.8 × 2 = 19.6 ms

time period = 19.6 ÷ 5 = 3.92 ms = 0.00392 s

f = 1 / T = 1 ÷ 0.00392 = 255 Hz (to 3 significant figures)

This is very close to the 256 Hz marked on the fork, so the measurement is accurate.

Compare this with using one wave. A single wave looks about 2.0 divisions wide, which gives T = 4.0 ms and f = 250 Hz. Measuring across 5 waves gives the better answer.

Results and checking

With a signal generator you can test several notes and compare each answer with the dial. You change the time base each time so that a few waves fill the screen.

Frequency on the dial in HzTime base in ms/divDivisions for one waveTime period in msMeasured frequency in Hz
12524.08.0125
32013.13.1323
8000.52.51.25800
10000.25.11.02980

The measured values are close to the dial values but not always exactly the same. Small differences come from reading the divisions on the screen, and the dial on the signal generator may not be perfectly accurate either.

Variables: when you test several notes, the frequency set on the signal generator is the independent variable. The time period measured from the trace is the dependent variable. Keep the same microphone, loudspeaker and oscilloscope throughout.

Safety: this is a low-risk practical. Keep the loudspeaker at a comfortable volume to protect your hearing. Strike the tuning fork on a cork or rubber block, never on a hard bench or near your eyes. The equipment runs from the mains, so keep liquids away and cables tidy.

Common mistakes

Forgetting to change milliseconds into seconds. Using T = 1.6 instead of 0.0016 gives a frequency that is 1000 times too small.

Measuring from a crest to the next trough. That is only half a wave. One complete wave is crest to crest.

Using the gain setting in the calculation. The gain is about the height of the trace. Only the time base is used to find the time period.

Counting crests instead of waves. Six crests in a row mark out five complete waves.

Leaving noise in the room. Other sounds reach the microphone and make the trace uneven, so work in a quiet room with the source close to the microphone.

Exam-style question

A student investigates the frequency of the sound from a tuning fork. She holds the vibrating fork near a microphone connected to an oscilloscope.

The trace shows 4 complete waves in 7.8 divisions across the screen. The time base setting is 1 ms/div.

(a) Describe how the student should adjust the oscilloscope to obtain a trace that is easy to measure. [2 marks]

(b) Calculate the frequency of the sound. Give your answer to 3 significant figures. [4 marks]

(c) Explain why the student measured across 4 waves instead of 1 wave. [2 marks]

(d) The tuning fork is marked 512 Hz. Comment on the student's result. [1 mark]

(e) Suggest one advantage of using a signal generator and loudspeaker instead of the tuning fork. [1 mark]

Model answer

(a) Adjust the gain so the trace is tall but still fits on the screen (1). Adjust the time base so the trace is steady and shows a few complete waves (1).

(b) time for 4 waves = 7.8 × 1 = 7.8 ms (1). time period = 7.8 ÷ 4 = 1.95 ms (1) = 0.00195 s (1). f = 1 / T = 1 ÷ 0.00195 = 513 Hz (1).

(c) The screen can only be read to a fraction of a division, and this reading error is shared between 4 waves (1), so the error in the time period of one wave is smaller and the frequency is more accurate (1).

(d) The result is accurate, because 513 Hz is very close to the 512 Hz marked on the fork (1).

(e) Any one from: the sound does not fade, so the trace stays steady; the note is louder, so the trace is taller; the frequency can be changed to test other notes (1).

Exam tip

Set the calculation out in three lines every time: divisions × time base, then milliseconds to seconds, then f = 1 / T. The examiner gives marks for each line, so you still score if a slip on the calculator spoils the final answer.

Test Your Knowledge